How to Perform a Pre-Cast SDS-PAGE
		
		
  
    Note: The upper stacking gel and sample buffer
    have a pH of 6.8 while the lower resolving gel has a pH of 8.8. The running
    buffer (a solution of Tris and glycine) is put in both buffer chambers
    and has a pH of 8.3.
  
		Assembling the Upper Buffer Chamber
These are the directions that come with the apparatus
and they read like directions of how to tie a shoe - not well. Nevertheless,
we need some written directions to insure a leakproof seal, make sure the
gray U-shaped inner cooling core gaskets are clean. Inspect the gasket for
small cuts that could result in an upper buffer leak. There are two sides
to this gasket. Make sure that the side with the notch is exposed for contact
with the gel sandwich.
Assembly 
Lay the inner cooling core flat on a lab bench.
With the glass plates of the gel sandwich facing the cooling core (and the
clamp screws facing out), carefully slide the clamp assembly wedges underneath
the locator slots on the inner cooling core until the inner glass plate
of the gel sandwich butts up against the notch in the U-shaped gasket.
Lubricating the raised portions of the U-shaped gasket with a drop of running
buffer or water helps the glass plate sandwich slide properly.
While pushing the clamp assembly slightly up toward
the top of the locator slots, snap the clamp assembly fully onto the cooling
core by pressing at the bottom of the clamp assembly until the cooling core
latch engages each side of the clamp assembly. (Do not pull out on cooling
core latch at the same time.) It is very important that there is good contact
at the point where the short glass plate contacts the notch in the gasket
to prevent upper buffer leaks. If the glass plate is not firmly seated in
its proper place against the notch in the gasket buffer leaks will occur.
Turn over the inner cooling core and attach another
clamp assembly to the other side of the core in the same manner.
Preparing Samples
  - Add 50 µL of bME 
    to 950 µL 3X loading dye.  
  
 - Heat to 95° C for 5 minutes. 
  
 - Load samples into wells once they have cooled to RT. 
     
 
Loading The Samples 
  - Prepare 300 ml of electrode buffer by combining
  30 ml of 10X electrode buffer with 270 ml of deionized water.
  
 - Lower the inner cooling core into the lower buffer
  chamber of the Mini-PROTEAN II cell. Add approximately 115 ml of buffer
  to the upper buffer chamber. Fill until the buffer reaches a level halfway
  between the short and long plates. Do not overfill the upper buffer chamber.
  Overfilling the upper buffer chamber will result in siphoning of the buffer
  over the top of the gasket, resulting in buffer loss and interruption of
  the electrophoresis.
  
 - Pour the remainder of the buffer into the lower
  buffer chamber so that at least the bottom 1 cm of the gel is covered.
  Remove any air bubbles from the bottom of the gel so that good electrical
  contact is achieved. This can be done by squirting the lower buffer with
  a curved pipette until the bubbles clear.
  
 - Load the samples into the wells under the electrode
  buffer with a pipette and one very thin tip. Insert the tip to about 1-2
  mm from the well bottom before delivery. 
 
Running The Gel
  - Place the lid on top of the lower buffer chamber
  to fully enclose the cell. the correct orientation is made by matching
  the colors of plugs on the lid with the jacks on the inner cooling core.
  
 - Attach the electrical leads to a suitable power
  supply (200 V minimum) with the proper polarity.
  
 - Apply power to the Mini-PROTEAN II cell and begin
  electrophoresis. The recommended power for optimal resolution with minimal
  thermal band distortion is 200 volts, constant voltage setting. No adjustment
  of the setting is necessary for spacer thickness, or number of gels. The
  usual run time is approximately 45 minutes. Current should be approximately
  60 mA per gel (120 mA for two gels) at the beginning of the run. During
  the 45 minute run the current will slowly drop to about 30 mA per gel.
  This drop is caused by the change in buffer ions in the gel, causing a
  slow rise in the resistance in the gel. 
 
Removing The Gel
  - After electrophoresis is complete, turn off the
  power supply and disconnect the electrical leads.
  
 - Remove the cell lid and carefully pull the inner
  cooling core out of the lower chamber. Pour off the upper buffer.
  
 - Lay the inner cooling core on its side and remove
  the gels and the attached clamp assembly.
  
 - Loosen all four screws of the clamp assembly
  and remove the gel.
  
 - Push one of the spacers of the gel out to the
  outside of the plates without removing it.
  
 - Gently twist the spacer so that the upper glass
  plate pulls away from the gel. Remove the plate. The gel will stick to
  one of the plates.
  
 - Float the gel off of the glass plate by inverting
  the gel and glass plate under fixative solution or blotting transfer buffer
  and agitate gently until the gel separates from the glass plate. 
 
4X sample buffer - store at RT°
  - 6.06 g Tris
  
 - 2 ml 20% SDS 
  
 - 20 ml glycerol
  
 - pH to 6.8 with HCl
  
 - water to 100 ml
  
 - 0.03% bromophenol blue
  
 - plus fresh DTT added just before use
 
		
		
Lab Schedule Outlined
Lab Schedule In Context of Research Project
		
			Molecular Biology Main Page
			Course Materials
		
		
			Biology Main Page
		
		
			
 
		
		
		© Copyright 2002 Department of Biology, Davidson College, Davidson, NC 28036
		Send comments, questions, and suggestions to: macampbell@davidson.edu